IL-26 CLIA Kit
-
- Target Alle IL-26 (IL26) CLIA Kits anzeigen
- IL-26 (IL26) (Interleukin 26 (IL26))
- Reaktivität
- Human
- Nachweismethode
- Chemiluminescent
- Methodentyp
- Sandwich ELISA
- Detektionsbereich
- 15.62 pg/mL - 1000 pg/mL
- Untere Nachweisgrenze
- 15.62 pg/mL
- Applikation
- ELISA
- Proben
- Cell Culture Supernatant, Cell Lysate, Plasma, Serum, Tissue Homogenate
- Analytische Methode
- Quantitative
- Spezifität
- This assay has high sensitivity and excellent specificity for detection of Interleukin 26 (IL26). No significant cross-reactivity or interference between Interleukin 26 (IL26) and analogues was observed.
- Sensitivität
- 0.51 pg/mL
- Featured
- Zu unserem meistverkauften IL26 ELISA Kit
- Top Product
- Discover our top product IL26 ELISA Kit
-
-
- Kommentare
-
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
- Testdauer
- 2 - 3 h
- Plattentyp
- Pre-coated
- Protokoll
- The microplate provided in this kit has been pre-coated with an antibody specific to Interleukin 26 (IL26). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Interleukin 26 (IL26). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Interleukin 26 (IL26) level in the sample or standard.,
- Testpräzision
-
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Interleukin 26 (IL26) were tested 20 times on one plate, respectively
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Interleukin 26 (IL26) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12% - Beschränkungen
- Nur für Forschungszwecke einsetzbar
-
- Handhabung
- Do not allow to contact skin or eyes. Calibrators, controls and specimen samples should be assayed in duplicate. Once the procedure has been started, all steps should be completed without interruption.
- Lagerung
- 4 °C,-20 °C
- Informationen zur Lagerung
- -20°C. Bring all reagents to room temperature before beginning test. The kit may be stored at 4°C for immediate use within two days upon arrival. Reseal any unused strips with desiccant pack. Minimize freeze/thaw cycles.
- Haltbarkeit
- 4-8 months
-
- Target Alle IL-26 (IL26) CLIA Kits anzeigen
- IL-26 (IL26) (Interleukin 26 (IL26))
- Andere Bezeichnung
- Interleukin 26 (IL26 Produkte)
- Synonyme
- AK155 CLIA Kit, IL-26 CLIA Kit, zgc:194942 CLIA Kit, IL26 CLIA Kit, interleukin 26 CLIA Kit, IL26 CLIA Kit, il26 CLIA Kit
- Hintergrund
-
Gene Name: Interleukin 26
Gene Aliases: AK155
- Gen-ID
- 55801
- UniProt
- Q9NPH9
-