CRP ELISA Kit
-
- Target Alle CRP ELISA Kits anzeigen
- CRP (C-Reactive Protein (CRP))
-
Reaktivität
- Human
- Nachweismethode
- Colorimetric
- Methodentyp
- Sandwich ELISA
- Detektionsbereich
- 62.5 pg/mL - 4000 pg/mL
- Untere Nachweisgrenze
- 62.5 pg/mL
- Applikation
- ELISA
- Verwendungszweck
-
The kit is a high sensitive sandwich enzyme immunoassay for in vitro quantitative measurement of CRP in human serum, plasma, tissue homogenates, cell lysates, urine, cerebrospinal fluid, cell culture supernates.
We offer validation data (WB) for the kit components. So you can be sure to order a reliable ELISA kit product composed of high quality reagents. - Proben
- Cell Culture Supernatant, Cell Lysate, Cerebrospinal Fluid, Plasma, Serum, Tissue Homogenate, Urine
- Analytische Methode
- Quantitative
- Spezifität
- This assay has high sensitivity and excellent specificity for detection of C Reactive Protein.
- Sensitivität
- 23.3 pg/mL
- Güteklasse
- High Sensitivity
- Bestandteile
-
- Pre-coated, ready to use 96-well strip plate, flat buttom
- Plate sealer for 96 wells
- Reference Standard
- Standard Diluent
- Detection Reagent A
- Detection Reagent B
- Assay Diluent A
- Assay Diluent B
- Reagent Diluent (if Detection Reagent is lyophilized)
- TMB Substrate
- Stop Solution
- Wash Buffer (30 x concentrate)
- Instruction manual
-
-
- Applikationshinweise
-
- Limited by the current condition and scientific technology, we cannot completely conduct the comprehensive identification and analysis on the raw material provided by suppliers. So there might be some qualitative and technical risks to use the kit.
- The final experimental results will be closely related to validity of the products, operation skills of the end users and the experimental environments. Please make sure that sufficient samples are available.
- Kits from different batches may be a little different in detection range, sensitivity and color developing time.
- Do not mix or substitute reagents from one kit lot to another. Use only the reagents supplied by manufacturer.
- Protect all reagents from strong light during storage and incubation. All the bottle caps of reagents should be covered tightly to prevent the evaporation and contamination of microorganism.
- There may be some foggy substance in the wells when the plate is opened at the first time. It will not have any effect on the final assay results. Do not remove microtiter plate from the storage bag until needed.
- Wrong operations during the reagents preparation and loading, as well as incorrect parameter setting for the plate reader may lead to incorrect results. A microplate plate reader with a bandwidth of 10nm or less and an optical density range of 0-3 O.D. or greater at 450 ± 10nm wavelength is acceptable for use in absorbance measurement. Please read the instruction carefully and adjust the instrument prior to the experiment.
- Even the same operator might get different results in two separate experiments. In order to get better reproducible results, the operation of every step in the assay should be controlled. Furthermore, a preliminary experiment before assay for each batch is recommended.
- Each kit has been strictly passed Q.C test. However, results from end users might be inconsistent with our in-house data due to some unexpected transportation conditions or different lab equipments. Intra-assay variance among kits from different batches might arise from above factors, too.
- Kits from different manufacturers for the same item might produce different results, since we have not compared our products with other manufacturers.
- Kommentare
-
Information on standard material:
The standard might be recombinant protein or natural protein, that will depend on the specific kit. Moreover, the expression system is E.coli or yeast or mammal cell. There is 0.05% proclin 300 in the standard as preservative.
Information on reagents:
The stop solution used in the kit is sulfuric acid with concentration of 1 mol/L. And the wash solution is TBS. The standard diluent contains 0.02 % sodium azide, assay diluent A and assay diluent B contain 0.01% sodium azide. Some kits can contain is BSA in them.
Information on antibodies:
The provided antibodies and their host vary in different kits. - Probenmenge
- 100 μL
- Testdauer
- 3 h
- Plattentyp
- Pre-coated
- Protokoll
-
- Prepare all reagents, samples and standards,
- Add 100μL standard or sample to each well. Incubate 1 hours at 37 °C,
- Aspirate and add 100μL prepared Detection Reagent A. Incubate 1 hour at 37 °C,
- Aspirate and wash 3 times,
- Add 100μL prepared Detection Reagent B. Incubate 30 minutes at 37 °C,
- Aspirate and wash 5 times,
- Add 90μL Substrate Solution. Incubate 10-20 minutes at 37 °C,
- Add 50μL Stop Solution. Read at 450nm immediately.
- Aufbereitung der Reagenzien
-
- Bring all kit components and samples to room temperature (18-25 °C) before use. If the kit is not used up all at once, remove only the strips and reagents for the current experiment and leave the remaining strips and reagents in the desired condition.
- Standard - Reconstitute the Standard with 1.0mL of Standard Diluent, kept for 10 minutes at room temperature, shake gently (not to foam). The concentration of the standard in the stock solution is 4,000 pg/mL. Prepare 7 tubes containing 0.5mL Standard Diluent and produce a double dilution series. Mix each tube thoroughly before the next transfer. Set up 7 points of diluted standard such as 4,000pg/mL, 2,000pg/mL, 1,000pg/mL, 500pg/mL, 250pg/mL, 125pg/mL, 62.5pg/mL, and the last tube with Standard Diluent is the blank as 0pg/mL.
- Detection Reagent A and Detection Reagent B - Spin or centrifuge the stock of Detection Reagent A and B briefly before use. Dilute to working concentration (1:100) with Assay Diluent A or B, respectively.
- Wash Solution - Dilute 20 mL of Wash Solution Concentrate (30x) with 580 mL of deionized or distilled water to make 600 mL of Wash Solution (1x).
- TMB Substrate - Aspirate the required amount of solution with sterile tip and do not return the residual solution back into the vial.
Note:
- Serial dilution directly in the wells is not recommended.
- Prepare standard within 15 minutes before assay. Do not dissolve the reagents directly at 37 °C.
- Detection Reagent A and B are sticky solutions, so pipette them slowly to reduce volume errors.
- Reconstitute Standard or working solutions of Detection Reagent A and B carefully according to instructions, avoiding foaming and mixing gently until crystals are completely dissolved. To minimize inaccuracy caused by pipetting, use small volumes and ensure pipettes are calibrated. It is recommended to aspirate more than 10 µL for one-time pipetting.
- The reconstituted Standard, Detection Reagent A and B can only be used once.
- When crystals have formed in the Wash Solution concentrate (30x), warm it to room temperature and mix gently until the crystals are completely dissolved.
- Contaminated water or preparation containers affect the detection result.
- Aufbereitung der Proben
-
- It is recommended to use fresh samples without long storage, otherwise protein degradation and denaturationmay occur in these samples, leading to false results. Samples should therefore be stored for a short periodat 2 - 8 °C or aliquoted at -20 °C (≤1 month) or -80 °C (≤ 3 months). Repeated freeze-thawcycles should be avoided. Prior to assay, the frozen samples should be slowly thawed and centrifuged toremove precipitates.
- If the sample type is not specified in the instructions, a preliminary test is necessary to determinecompatibility with the kit.
- If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibilityof causing a deviation due to the introduced chemical substance.The recommended dilution factor is for reference only.
- Please estimate the concentration of the samples before performing the test. If the values are not in therange of the standard curve, the optimal sample dilution for the particular experiment has to be determined.Samples should then be diluted with PBS (pH =7.0-7.2).
- Testpräzision
-
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level of target were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level of target were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV < 10%
Inter-Assay: CV < 12% - Beschränkungen
- Nur für Forschungszwecke einsetzbar
-
- Vorsichtsmaßnahmen
- The Stop Solution suggested for use with this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.
- Handhabung
-
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end. - Lagerung
- 4 °C/-20 °C
- Informationen zur Lagerung
-
- For unopened kit: All reagents should be stored according to the labels on the vials. The Standard, Detection Reagent A, Detection Reagent B, and 96-well Strip Plate should be stored at -20 °C upon receipt, while the other reagents should be stored at 4 °C.
- For opened kits: the remaining reagents must be stored according to the above storage conditions. In addition, please return the unused wells to the foil pouch containing the desiccant and seal the foil pouch with the zipper.
- Haltbarkeit
- 6 months
-
-
The effects of non-surgical periodontal treatment on glycemic control, oxidative stress balance and quality of life in patients with type 2 diabetes: A randomized clinical trial." in: PLoS ONE, Vol. 12, Issue 11, pp. e0188171, (2017) (PubMed).
: "Anti-atherosclerosis and cardio-protective effects of the Angong Niuhuang Pill on a high fat and vitamin D3 induced rodent model of atherosclerosis." in: Journal of ethnopharmacology, Vol. 195, pp. 118-126, (2016) (PubMed).
: "Among overweight middle-aged men, first-borns have lower insulin sensitivity than second-borns." in: Scientific reports, Vol. 4, pp. 3906, (2014) (PubMed).
: "Olive (Olea europaea L.) leaf polyphenols improve insulin sensitivity in middle-aged overweight men: a randomized, placebo-controlled, crossover trial." in: PLoS ONE, Vol. 8, Issue 3, pp. e57622, (2013) (PubMed).
: "Relationships among serum C-reactive protein, receptor for advanced glycation products, metabolic dysfunction, and cognitive impairments." in: BMC neurology, Vol. 13, pp. 110, (2013) (PubMed).
: "Effect of high free fatty acids on the anti-contractile response of perivascular adipose tissue in rat aorta." in: Journal of molecular and cellular cardiology, Vol. 63, pp. 169-74, (2013) (PubMed).
: "[Replacement of the upper arm flexor muscles by m. latissimus dorsi in partial paralysis of the arm plexus]." in: Handchirurgie, Mikrochirurgie, plastische Chirurgie : Organ der Deutschsprachigen Arbeitsgemeinschaft fur Handchirurgie : Organ der Deutschsprachigen Arbeitsgemeinschaft fur Mikrochirurgie der Peripheren Nerven und Gefasse : Organ der V..., Vol. 19, Issue 3, pp. 171-3, (1987) (PubMed).
: "[Interaction of recovery cycles of evoked potentials in the cerebral cortex of the cat in response to stimulation of the superior colliculi and pulvinar]." in: Zhurnal vyssheĭ nervnoĭ deiatelnosti imeni I P Pavlova, Vol. 36, Issue 4, pp. 715-20, (1986) (PubMed).
: "
-
The effects of non-surgical periodontal treatment on glycemic control, oxidative stress balance and quality of life in patients with type 2 diabetes: A randomized clinical trial." in: PLoS ONE, Vol. 12, Issue 11, pp. e0188171, (2017) (PubMed).
-
- Target Alle CRP ELISA Kits anzeigen
- CRP (C-Reactive Protein (CRP))
- Andere Bezeichnung
- C Reactive Protein (CRP) (CRP Produkte)
- Synonyme
- PTX1 ELISA Kit, crp ELISA Kit, AI255847 ELISA Kit, Aa1249 ELISA Kit, Ab1-341 ELISA Kit, Ab2-196 ELISA Kit, Ac1-114 ELISA Kit, Ac1262 ELISA Kit, Ac2-069 ELISA Kit, Ba2-693 ELISA Kit, APCS ELISA Kit, 0610010I23Rik ELISA Kit, AW743261 ELISA Kit, C77570 ELISA Kit, CRP2 ELISA Kit, CRP4 ELISA Kit, Crp ELISA Kit, ESP1 ELISA Kit, Hlp ELISA Kit, CRP5.1 ELISA Kit, zgc:152809 ELISA Kit, C-reactive protein ELISA Kit, C-reactive protein, pentraxin-related ELISA Kit, c-reactive protein, pentraxin-related ELISA Kit, cysteine rich protein 2 ELISA Kit, CRP ELISA Kit, crp ELISA Kit, Crp ELISA Kit, Crip2 ELISA Kit, LOC776376 ELISA Kit
- Hintergrund
- C-RP, PTX1, Pentraxin-Related, Pentraxin 1
- UniProt
- P02741
- Pathways
- Carbohydrate Homeostasis
-