COVID-19 (SARS-CoV-2) N-Protein Human Antibody Detection Kit (Indirect ELISA method) for the semi-quantitative measurement of human IgM antibody against SARS-CoV-2 in serum or plasma.
Proben
Plasma, Serum
Analytische Methode
Semi-Quantitative
Produktmerkmale
COVID-19 N-Protein Human IgM ELISA Kit
Bestandteile
SARS-CoV-2 N-protein coated 96-well microplate
Wash Buffer
Stop Solution
Assay Diluent
Positive Control
HRP-anti positive control IgG Antibody
TMB One-Step Substrate
Benötigtes Material
Distilled or deionized water
Precision pipettes to deliver 2 μL to 1 μL volumes
Adjustable 1-25 μL pipettes for reagent preparation
100 μL and 1 liter graduated cylinders
Tubes to prepare standard and sample dilutions
Absorbent paper
Microplate reader capable of measuring absorbance at 450nm
Log-log graph paper or computer and software for ELISA data analysis
This COVID19 human IgM antibody ELISA kit employs an indirect ELISA method. In this kit, standard 96-well plates (12 strips with 8 wells/strip) are coated with the SARSCoV-2 N protein, which combines with the corresponding antibody present in a sample. When a secondary anti-human Antibody-HRP is added, a complex of Antibody-HRPhuman IgM antibody-virus N antigen forms on the microplate. A TMB substrate is added and a blue color is generated. The depth of color is relative to the amount of the anti-SARS-CoV-2 IgM antibody present. The Stop Solution changes the color from blue to yellow, and the intensity of the color is measured at 450 nm.
Testdurchführung
Prepare all reagents, samples and standards as instructed.
Add 100 μL positive control or sample to each well. Incubate 1.5 hours at room temperature.
Add 100 μL prepared HRP- anti positive control IgG Antibody and/or 100 μL of prepared HRP-anti Human IgG Antibody to each well. Incubate 1 hour at room temperature.
Add 100 μL TMB One-Step Substrate Reagent to each well. Incubate 10 minutes at room temperature.
Add 50 μL Stop Solution to each well. Read at 450 nm immediately.
Read at 450 nm immediately.
Beschränkungen
Nur für Forschungszwecke einsetzbar
Lagerung
-20 °C
Informationen zur Lagerung
The entire kit may be stored at -20°C for up to 1 year from the date of shipment. Avoid repeated freeze-thaw cycles. The kit may be stored at 4°C for up to 6 months. For extended storage, it is recommended to store at -80°C.