Telefon:
+49 (0)241 95 163 153
Fax:
+49 (0)241 95 163 155
E-Mail:
orders@antikoerper-online.de

FLT1 ELISA Kit

FLT1 Reaktivität: Human Colorimetric Sandwich ELISA 125 pg/mL - 8000 pg/mL Cell Culture Supernatant, Plasma, Serum
Produktnummer ABIN6963502
  • Target Alle FLT1 ELISA Kits anzeigen
    FLT1 (Fms-Related tyrosine Kinase 1 (VEGFR1) (FLT1))
    Reaktivität
    • 10
    • 8
    • 5
    • 3
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Human
    Nachweismethode
    Colorimetric
    Methodentyp
    Sandwich ELISA
    Detektionsbereich
    125 pg/mL - 8000 pg/mL
    Untere Nachweisgrenze
    125 pg/mL
    Applikation
    ELISA
    Verwendungszweck
    The kit is a sandwich enzyme immunoassay technique for the in vitro quantitative measurement in various sample types.
    Proben
    Cell Culture Supernatant, Plasma, Serum
    Analytische Methode
    Quantitative
    Spezifität
    This kit recognizes Human VEGFR1/FLT1 in samples. No Significant cross-reactivity or interference between Human VEGFR1/FLT1 and analogues was observed.
    Sensitivität
    75 pg/mL
    Bestandteile
    • Pre-coated, ready to use 96-well strip plate, flat buttom
    • Plate sealer for 96 wells
    • Reference Standard
    • Reference Standard & Sample Diluent
    • Biotinylated Detection Antibody (100 x concentrate)
    • HRP Conjugate (100 x concentrate)
    • Biotinylated Detection Antibody Diluent
    • HRP Conjugate Diluent
    • Substrate Reagent
    • Stop Solution
    • Wash Buffer (25 x concentrate)
    • Instruction manual
    Featured
    Zu unserem meistverkauften FLT1 ELISA Kit
  • Probenmenge
    100 μL
    Testdauer
    3.5 h
    Plattentyp
    Pre-coated
    Protokoll
    1. Add 100 µL standard or sample to each well. Incubate for 90 min at 37 °C.
    2. Remove the liquid. Add 100 µL Biotinylated Detection Antibody. Incubate for 1 hour at 37 °C.
    3. Aspirate and wash 3 times.
    4. Add 100 µL HRP Conjugate. Incubate for 30 min at 37 °C.
    5. Aspirate and wash 5 times.
    6. Add 90 µL Substrate Reagent. Incubate for 15 min at 37 °C.
    7. Add 50 µL Stop Solution. Read at 450 nm immediately.
    8. Calculation of results.
    Aufbereitung der Reagenzien
    1. Bring all reagents to room temperature (18~25 °C) before use. Follow the Microplate reader manual for set-up and preheat it for 15 min before OD measurement.
    2. Wash Buffer: Dilute 30 mL of Concentrated Wash Buffer with 720 mL of deionized or distilled water to prepare 750 mL of Wash Buffer.Note: if crystals have formed in the concentrate, warm it in a 40 °C water bath and mix it gently until the crystals have completely dissolved
    3. Standard working solution: Centrifuge the standard at 10,000xg for 1 min. Add 1.0 mL of Reference Standard &Sample Diluent, let it stand for 10 min and invert it gently several times. After it dissolves fully, mix it thoroughly with a pipette. This reconstitution produces a working solution of 8000 pg/mL. Then make serial dilutions as needed. The recommended dilution gradient is as follows: 8000, 4000, 2000, 1000, 500, 250, 125.00, 0 pg/mL. Dilution method: Take 7 EP tubes, add 500 μLof Reference Standard & Sample Diluent to each tube. Pipette 500 μLof the 8000 pg/mL working solution to the first tube and mix up to produce a 4000 pg/mL working solution. Pipette 500 μLof the solution from the former tube into the latter one according to these steps. The illustration below is for reference. Note: the last tube is regarded as a blank. Don't pipette solution into it from the former tube.
    4. Biotinylated Detection Antibody working solution: Calculate the required amount before the experiment (100 μL/well). In preparation, slightly more than calculated should be prepared. Centrifuge the stock tube before use, dilute the 100x Concentrated Biotinylated Detection Antibody to 1xworking solution with Biotinylated Detection Antibody Diluent.
    5. Concentrated HRP Conjugate working solution: Calculate the required amount before the experiment (100 μL/well). In preparation, slightly more than calculated should be prepared. Dilute the 100x Concentrated HRP Conjugate to 1x working solution with Concentrated HRP Conjugate Diluent.
    Aufbereitung der Proben
    • It is recommended to use fresh samples without long storage, otherwise protein degradation and denaturation may occur in these samples, leading to false results. Samples should therefore be stored for a short period at 2 - 8 °C or aliquoted at -20 °C (≤1 month) or -80 °C (≤ 3 months). Repeated freeze-thaw cycles should be avoided. Prior to assay, the frozen samples should be slowly thawed and centrifuged to remove precipitates.
    • If the sample type is not specified in the instructions, a preliminary test is necessary to determine compatibility with the kit.
    • If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibility of causing a deviation due to the introduced chemical substance. The recommended dilution factor is for reference only.
    • Please estimate the concentration of the samples before performing the test. If the values are not in the range of the standard curve, the optimal sample dilution for the particular experiment has to be determined.
    Testpräzision
    Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human VEGFR1/FLT1 were tested 20 times on one plate, respectively.
    Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human VEGFR1/FLT1 were tested on 3 different plates, 20 replicates in each plate.
    Both intra-CV and inter-CV are < 10 %.
    Beschränkungen
    Nur für Forschungszwecke einsetzbar
  • Lagerung
    4 °C,-20 °C
    Informationen zur Lagerung
    1. For unopened kit: All reagents should be stored according to the labels on the vials, so they are stable up to 12 months after receipt of the kit. The Reference Standard, Biotinylated Detection Antibody, HRP Conjugate and the 96-well stripe plate should be stored at -20 °C upon receipt while the other reagents should be stored at 4 °C.
    2. For used kit: When the kit is used, the remaining reagents need to be stored according to the above storage condition. Besides, please return the unused wells to the foil pouch containing the desiccant pack, and zip-seal the foil pouch.
    Haltbarkeit
    12 months
  • Uyanıkoğlu, İncebıyık, Turp, Çakmak, Sak, Hilali: "Serum Angiogenic and Anti-angiogenic Markers in Pregnant Women with Placenta Percreta." in: Balkan medical journal, Vol. 35, Issue 1, pp. 55-60, (2018) (PubMed).

  • Target Alle FLT1 ELISA Kits anzeigen
    FLT1 (Fms-Related tyrosine Kinase 1 (VEGFR1) (FLT1))
    Andere Bezeichnung
    Vascular Endothelial Growth Factor Receptor 1 (FLT1 Produkte)
    Synonyme
    FLT ELISA Kit, FLT-1 ELISA Kit, VEGFR-1 ELISA Kit, VEGFR1 ELISA Kit, FLT1 ELISA Kit, flt1b ELISA Kit, sFlt1 ELISA Kit, AI323757 ELISA Kit, Flt-1 ELISA Kit, fms related tyrosine kinase 1 ELISA Kit, FMS-related tyrosine kinase 1 ELISA Kit, vascular endothelial growth factor receptor 1 ELISA Kit, fms-related tyrosine kinase 1 (vascular endothelial growth factor/vascular permeability factor receptor) ELISA Kit, FMS-like tyrosine kinase 1 ELISA Kit, vascular endothelial growth factor receptor-1 ELISA Kit, FLT1 ELISA Kit, Flt1 ELISA Kit, CpipJ_CPIJ012087 ELISA Kit, flt1 ELISA Kit, FLT-1 ELISA Kit
    Hintergrund
    FLT, FLT-1, VEGFR-1, Fms-related tyrosine kinase 1
    Pathways
    RTK Signalweg, Signaling Events mediated by VEGFR1 and VEGFR2, VEGFR1 Specific Signals
Sie sind hier:
Kundenservice