TNNI3 ELISA Kit
-
- Target Alle TNNI3 ELISA Kits anzeigen
- TNNI3 (Cardiac Troponin I (TNNI3))
-
Reaktivität
- Human
- Nachweismethode
- Colorimetric
- Methodentyp
- Sandwich ELISA
- Detektionsbereich
- 47 pg/mL - 3000 pg/mL
- Untere Nachweisgrenze
- 47 pg/mL
- Applikation
- ELISA
- Verwendungszweck
- For the quantitative determination of human cardiac troponin I (cTn-I) concentrations in serum, plasma, urine.
- Proben
- Plasma, Serum, Urine
- Analytische Methode
- Quantitative
- Spezifität
- This assay has high sensitivity and excellent specificity for detection of human cTn-I. No significant cross-reactivity or interference between human cTn-I and analogues was observed. Note: Limited by current skills and knowledge, it is impossible for us to complete the cross-reactivity detection between human cTn-I and all the analogues, therefore, cross reaction may still exist.
- Sensitivität
- 11.75 pg/mL
- Bestandteile
-
- Assay plate
- Standard
- HRP-avidin (100 x concentrate)
- Biotin-antibody (100 x concentrate)
- Sample Diluent
- HRP-avidin Diluent
- Biotin-antibody Diluent
- Wash Buffer (25 x concentrate)
- TMB Substrate
- Stop Solution
- Adhesive Strip
- Top Product
- Discover our top product TNNI3 ELISA Kit
-
-
- Applikationshinweise
- Optimal working dilution should be determined by the investigator.
- Probenmenge
- 100 μL
- Testdauer
- 1 - 4.5 h
- Plattentyp
- Pre-coated
- Protokoll
-
- Prepare reagents, samples and standards as instructed.
- Add 100 µL standard or sample to each well. Incubate 2 hours at 37 °C.
- Remove the liquid of each well, don't wash.
- Add 100 µL Biotin-antibody (1x) to each well. Incubate 1 hour at 37 °C.
- Aspirate and wash 3 times.
- Add 100 µL HRP-avidin (1x) to each well. Incubate 1 hour at 37 °C
- Aspirate and wash 5 times.
- Add 90 μL of TMB Substrate to each well. Incubate for 15-30 minutes at 37 °C. Protect from light.
- Add 50 µL Stop Solution to each well. Read at 450 nm within 5 minutes.
- Aufbereitung der Reagenzien
-
- Biotin-antibody (1x) - Centrifuge the vial before opening. Biotin-antibody requires a 100-fold dilution. A suggested 100-fold dilution is 10 μL of Biotin-antibody + 990 μL of Biotin-antibody Diluent.
- HRP-avidin (1x) - Centrifuge the vial before opening. HRP-avidin requires a 100-fold dilution. A suggested 100-fold dilution is 10 μL of HRP-avidin + 990 μL of HRP-avidin Diluent.
- Wash Buffer(1x)- If crystals have formed in the concentrate, warm up to room temperature and mix gently until the crystals have completely dissolved. Dilute 20 mL of Wash Buffer Concentrate (25 x) into deionized or distilled water to prepare 500 mL of Wash Buffer (1 x).
- Standard Centrifuge the standard vial at 6000-10000rpm for 30s. Reconstitute the Standard with 1.0 mL of Sample Diluent. Do not substitute other diluents. This reconstitution produces a stock solution of 3000 pg/mL. Mix the standard to ensure complete reconstitution and allow the standard to sit for a minimum of 15 minutes with gentle agitation prior to making dilutions. Pipette 250 μL of Sample Diluent into each tube (S0-S6). Use the stock solution to produce a 2-fold dilution series (below). Mix each tube thoroughly before the next transfer. The undiluted Standard serves as the high standard (3000 pg/mL). Sample Diluent serves as the zero standard (0 pg/mL).
- Kindly use graduated containers to prepare the reagent. Please don't prepare the reagent directly in the Diluent vials provided in the kit.
- Bring all reagents to room temperature (18-25 °C) before use for 30 min.
- Prepare fresh standard for each assay. Use within 4 hours and discard after use.
- Making serial dilution in the wells directly is not permitted.
- Please carefully reconstitute Standards according to the instruction, and avoid foaming and mix gently until the crystals have completely dissolved. To minimize imprecision caused by pipetting, use small volumes and ensure that pipettors are calibrated. It is recommended to suck more than 10 μL for once pipetting.
- Distilled water is recommended to be used to make the preparation for reagents. Contaminated water or container for reagent preparation will influence the detection result.
- Aufbereitung der Proben
-
- It is recommended to use fresh samples without long storage, otherwise protein degradation and denaturationmay occur in these samples, leading to false results. Samples should therefore be stored for a short periodat 2 - 8 °C or aliquoted at -20 °C (≤1 month) or -80 °C (≤ 3 months). Repeated freeze-thawcycles should be avoided. Prior to assay, the frozen samples should be slowly thawed and centrifuged toremove precipitates.
- If the sample type is not specified in the instructions, a preliminary test is necessary to determinecompatibility with the kit.
- If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibilityof causing a deviation due to the introduced chemical substance.The recommended dilution factor is for reference only.
- Please estimate the concentration of the samples before performing the test. If the values are not in therange of the standard curve, the optimal sample dilution for the particular experiment has to be determined.Samples should then be diluted with PBS (pH =7.0-7.2).
- Testpräzision
-
Intra-assay Precision (Precision within an assay): CV%<8% Three samples of known concentration were tested twenty times on one plate to assess.
Inter-assay Precision (Precision between assays): CV%<10% Three samples of known concentration were tested in twenty assays to assess. - Beschränkungen
- Nur für Forschungszwecke einsetzbar
-
- Lagerung
- 4 °C,-20 °C
- Informationen zur Lagerung
- Unopened kit Store at 2 - 8°C. Do not use the kit beyond the expiration date May be stored for up to 1 month at 2 - 8°C. Coated assay Try to keep it in a sealed aluminum foil bag, plate and avoid the damp. Standard May be stored for up to 1 month at 2 - 8° C. If Biotin-antibody don't make recent use, better keep it store at HRP-avidin -20°C. Biotin-antibody Diluent Opened kit HRP-avidin Diluent Sample May be stored for up to 1 month at 2 - 8°C. Diluent Wash Buffer TMB Substrate Stop Solution *Provided this is within the expiration date of the kit.
- Haltbarkeit
- 6 months
-
- Target Alle TNNI3 ELISA Kits anzeigen
- TNNI3 (Cardiac Troponin I (TNNI3))
- Andere Bezeichnung
- troponin I type 3 (cardiac) (TNNI3 Produkte)
- Synonyme
- CMD1FF ELISA Kit, CMD2A ELISA Kit, CMH7 ELISA Kit, RCM1 ELISA Kit, TNNC1 ELISA Kit, cTnI ELISA Kit, CMD1Z ELISA Kit, CMH13 ELISA Kit, TN-C ELISA Kit, TNC ELISA Kit, TNNC ELISA Kit, Tn1 ELISA Kit, TnIc ELISA Kit, ctnIc ELISA Kit, XTnIc ELISA Kit, c-troponin ELISA Kit, cTNT ELISA Kit, cmh7 ELISA Kit, ctni ELISA Kit, tnnc1 ELISA Kit, TNNI3 ELISA Kit, TnI ELISA Kit, cTNI ELISA Kit, AI874626 ELISA Kit, TnC ELISA Kit, cTnC ELISA Kit, tncc ELISA Kit, Tncc ELISA Kit, troponin I3, cardiac type ELISA Kit, troponin C1, slow skeletal and cardiac type ELISA Kit, troponin I, cardiac 3 ELISA Kit, troponin I type 3 (cardiac) ELISA Kit, troponin I3, cardiac type S homeolog ELISA Kit, troponin I3, cardiac type L homeolog ELISA Kit, cardiac troponin I ELISA Kit, troponin C, cardiac/slow skeletal ELISA Kit, TNNI3 ELISA Kit, TNNC1 ELISA Kit, Tnni3 ELISA Kit, tnni3.S ELISA Kit, tnni3.L ELISA Kit, tnni3 ELISA Kit, LOC100462680 ELISA Kit, Tnnc1 ELISA Kit
- Hintergrund
-
Abbreviation: TNNI3
Alias: CMD2A, CMH7, MGC116817, RCM1, TNNC1, cTnI, familial hypertrophic cardiomyopathy 7|troponin I, cardiac
- UniProt
- P19429
-