APOA1 ELISA Kit

APOA1 Reaktivität: Rind (Kuh) Colorimetric Competition ELISA 78 ng/mL - 5000 ng/mL Plasma, Serum, Tissue Homogenate
Produktnummer ABIN6976153
  • Target Alle APOA1 ELISA Kits anzeigen
    APOA1 (Apolipoprotein A-I (APOA1))
    Reaktivität
    • 13
    • 8
    • 5
    • 5
    • 5
    • 4
    • 3
    • 2
    • 2
    • 2
    • 1
    • 1
    • 1
    Rind (Kuh)
    Nachweismethode
    Colorimetric
    Methodentyp
    Competition ELISA
    Detektionsbereich
    78 ng/mL - 5000 ng/mL
    Untere Nachweisgrenze
    78 ng/mL
    Applikation
    ELISA
    Verwendungszweck
    For the quantitative determination of bovine apolipoprotein A-I (APOA1) concentrations in serum, plasma, tissue homogenates.
    Proben
    Plasma, Serum, Tissue Homogenate
    Analytische Methode
    Quantitative
    Spezifität
    This assay has high sensitivity and excellent specificity for detection of bovine APOA1. No significant cross-reactivity or interference between bovine APOA1 and analogues was observed. Note: Limited by current skills and knowledge, it is impossible for us to complete the cross-reactivity detection between bovine APOA1 and all the analogues, therefore, cross reaction may still exist.
    Sensitivität
    19.5 ng/mL
    Bestandteile
    • Assay plate
    • Standard
    • HRP-conjugate (100 x concentrate)
    • Sample Diluent
    • HRP-conjugate Diluent
    • Wash Buffer (25 x concentrate)
    • TMB Substrate
    • Stop Solution
    • Adhesive Strip
  • Stop Solution
  • Adhesive Strip
  • Top Product
    Discover our top product APOA1 ELISA Kit
  • Applikationshinweise
    Optimal working dilution should be determined by the investigator.
    Probenmenge
    50 μL
    Testdauer
    1 - 4.5 h
    Plattentyp
    Pre-coated
    Protokoll
    1. Prepare reagents, samples and standards as instructed.
    2. Set a Blank well without any solution.
    3. Add 50 µL standard or sample to each well.
    4. Add 50 µL HRP-conjugate (1x) to each well (Not to Blank well).
    5. Incubate 1 hour at 37 °C
    6. Aspirate and wash 5 times.
    7. Add 90 μL of TMB Substrate to each well. Incubate for 20 minutes at 37 °C. Protect from light.
    8. Add 50 µL Stop Solution to each well. Read at 450 nm within 5 minutes.
    Aufbereitung der Reagenzien
    1. HRP-conjugate (1x) - Centrifuge the vial before opening. HRP-conjugate requires a 100-fold dilution. A suggested 100-fold dilution is 10 μL of HRP-conjugate + 990 μL of HRP-conjugate Diluent.
    2. Wash Buffer(1x)- If crystals have formed in the concentrate, warm up to room temperature and mix gently until the crystals have completely dissolved. Dilute 20 mL of Wash Buffer Concentrate (25 x) into deionized or distilled water to prepare 500 mL of Wash Buffer (1 x).
    3. Standard Centrifuge the standard vial at 6000-10000rpm for 30s. Reconstitute the Standard with 1.0 mL of Sample Diluent. Do not substitute other diluents. This reconstitution produces a stock solution of 5000 ng/mL. Mix the standard to ensure complete reconstitution and allow the standard to sit for a minimum of 15 minutes with gentle agitation prior to making dilutions. Pipette 150 μL of Sample Diluent into each tube (S0-S6). Use the stock solution to produce a 2-fold dilution series (below). Mix each tube thoroughly before the next transfer. The undiluted Standard serves as the high standard (5000 ng/mL). Sample Diluent serves as the zero standard (0 ng/mL).
    Note:
    • Kindly use graduated containers to prepare the reagent. Please don't prepare the reagent directly in the Diluent vials provided in the kit.
    • Bring all reagents to room temperature (18-25 °C) before use for 30 min.
    • Prepare fresh standard for each assay. Use within 4 hours and discard after use.
    • Making serial dilution in the wells directly is not permitted.
    • Please carefully reconstitute Standards according to the instruction, and avoid foaming and mix gently until the crystals have completely dissolved. To minimize imprecision caused by pipetting, use small volumes and ensure that pipettors are calibrated. It is recommended to suck more than 10 μL for once pipetting.
    • Distilled water is recommended to be used to make the preparation for reagents. Contaminated water or container for reagent preparation will influence the detection result.
    Aufbereitung der Proben
    • It is recommended to use fresh samples without long storage, otherwise protein degradation and denaturationmay occur in these samples, leading to false results. Samples should therefore be stored for a short periodat 2 - 8 °C or aliquoted at -20 °C (≤1 month) or -80 °C (≤ 3 months). Repeated freeze-thawcycles should be avoided. Prior to assay, the frozen samples should be slowly thawed and centrifuged toremove precipitates.
    • If the sample type is not specified in the instructions, a preliminary test is necessary to determinecompatibility with the kit.
    • If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibilityof causing a deviation due to the introduced chemical substance.The recommended dilution factor is for reference only.
    • Please estimate the concentration of the samples before performing the test. If the values are not in therange of the standard curve, the optimal sample dilution for the particular experiment has to be determined.Samples should then be diluted with PBS (pH =7.0-7.2).
    Note:
    Recommend to dilute the serum or plasma samples with Sample Diluent (1:2000) before test. The suggested 2000-fold dilution can be achieved by adding 2 μL sample to 98 μL of Sample Diluent. Complete the 2000-fold dilution by adding 6 μL of this solution to 234 μL of Sample Diluent. The recommended dilution factor is for reference only. The optimal dilution factor should be determined by users according to their particular experiments. 6
    Testpräzision
    Intra-assay Precision (Precision within an assay): CV%<8% Three samples of known concentration were tested twenty times on one plate to assess.
    Inter-assay Precision (Precision between assays): CV%<10% Three samples of known concentration were tested in twenty assays to assess.
    Beschränkungen
    Nur für Forschungszwecke einsetzbar
  • Lagerung
    4 °C,-20 °C
    Informationen zur Lagerung
    Unopened kit Store at 2 - 8°C. Do not use the kit beyond the expiration date. May be stored for up to 1 month at 2 - 8°C. Coated assay Try to keep it in a sealed aluminum foil bag, plate and avoid the damp. Standard May be stored for up to 1 month at 2 - 8° C. If don't make recent use, better keep it store at HRP-conjugate -20°C. Opened kit HRP-conjugate Diluent Sample Diluent May be stored for up to 1 month at 2 - 8°C. Wash Buffer TMB Substrate Stop Solution *Provided this is within the expiration date of the kit.
    Haltbarkeit
    6 months
  • Target Alle APOA1 ELISA Kits anzeigen
    APOA1 (Apolipoprotein A-I (APOA1))
    Andere Bezeichnung
    apolipoprotein A-I (APOA1 Produkte)
    Synonyme
    Alp-1 ELISA Kit, Apoa-1 ELISA Kit, Brp-14 ELISA Kit, Ltw-1 ELISA Kit, Lvtw-1 ELISA Kit, Sep-1 ELISA Kit, Sep-2 ELISA Kit, Sep2 ELISA Kit, apoA-I ELISA Kit, apoa ELISA Kit, apoa1 ELISA Kit, cb49 ELISA Kit, wu:fb33f01 ELISA Kit, zgc:103718 ELISA Kit, MGC64335 ELISA Kit, MGC89745 ELISA Kit, Apo-AI ELISA Kit, ApoA-I ELISA Kit, APOA1 ELISA Kit, LOC100136573 ELISA Kit, apoa-i ELISA Kit, ApoAI ELISA Kit, apolipoprotein A1 ELISA Kit, apolipoprotein A-I ELISA Kit, apolipoprotein A-Ia ELISA Kit, apolipoprotein A-I S homeolog ELISA Kit, apolipoprotein A4 ELISA Kit, APOA1 ELISA Kit, Apoa1 ELISA Kit, apoa1a ELISA Kit, apoa1.S ELISA Kit, apoa1 ELISA Kit, LOC100136573 ELISA Kit, apoa-i ELISA Kit, apoa4 ELISA Kit
    Hintergrund

    Abbreviation: APOA1

    Alias: MGC117399, OTTHUMP00000069346|OTTHUMP00000069347|OTTHUMP00000069348

    UniProt
    P15497
    Pathways
    Regulation of Lipid Metabolism by PPARalpha, Production of Molecular Mediator of Immune Response, Lipid Metabolism
Sie sind hier:
Kundenservice