APOA4 ELISA Kit

High Sensitivity APOA4 Reaktivität: Maus Colorimetric Sandwich ELISA 0.156 ng/mL - 10 ng/mL Plasma, Serum
Produktnummer ABIN6999458
  • Target Alle APOA4 ELISA Kits anzeigen
    APOA4 (Apolipoprotein A-IV (APOA4))
    Reaktivität
    • 7
    • 5
    • 4
    • 3
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Maus
    Nachweismethode
    Colorimetric
    Methodentyp
    Sandwich ELISA
    Detektionsbereich
    0.156 ng/mL - 10 ng/mL
    Untere Nachweisgrenze
    0.156 ng/mL
    Applikation
    ELISA
    Verwendungszweck
    The kit is a high sensitivity sandwich enzyme immunoassay for in vitro quantitative measurement in various sample types.
    Proben
    Plasma, Serum
    Analytische Methode
    Quantitative
    Spezifität
    This assay has high sensitivity and excellent specificity for detection of Apolipoprotein A4.
    Sensitivität
    0.064 ng/mL
    Güteklasse
    High Sensitivity
    Bestandteile
    • Pre-coated, ready to use 96-well strip plate, flat buttom
    • Plate sealer for 96 wells
    • Reference Standard
    • Standard Diluent
    • Detection Reagent A
    • Detection Reagent B
    • Assay Diluent A
    • Assay Diluent B
    • Reagent Diluent (if Detection Reagent is lyophilized)
    • TMB Substrate
    • Stop Solution
    • Wash Buffer (30 x concentrate)
    • Instruction manual
    Top Product
    Discover our top product APOA4 ELISA Kit
  • Probenmenge
    100 µL
    Testdauer
    3 h
    Plattentyp
    Pre-coated
    Protokoll
    1. Prepare all reagents, samples and standards,
    2. Add 100μL standard or sample to each well. Incubate 1 hours at 37 °C,
    3. Aspirate and add 100μL prepared Detection Reagent A. Incubate 1 hour at 37 °C,
    4. Aspirate and wash 3 times,
    5. Add 100μL prepared Detection Reagent B. Incubate 30 minutes at 37 °C,
    6. Aspirate and wash 5 times,
    7. Add 90μL Substrate Solution. Incubate 10-20 minutes at 37 °C,
    8. Add 50μL Stop Solution. Read at 450nm immediately.
    Aufbereitung der Reagenzien
    1. Bring all kit components and samples to room temperature (18-25 °C) before use. If the kit is not used up all at once, remove only the strips and reagents for the current experiment and leave the remaining strips and reagents in the desired condition.
    2. Standard - Reconstitute the Standard with 0.5mL of Standard Diluent, kept for 10 minutes at room temperature, shake gently (not to foam). The concentration of the standard in the stock solution is 1025ng/mL. Prepare 7 tubes containing 0.25mL Standard Diluent and produce a double dilution series. Mix each tube thoroughly before the next transfer. Set up 7 points of diluted standard such as 10ng/mL, 5ng/mL, 2.5ng/mL, 1.25ng/mL, 0.625ng/mL, 0.312ng/mL, 0.156ng/mL, and the last tubes with Standard Diluent is the blank as 0ng/mL.
    3. Detection Reagent A and Detection Reagent B - Spin or centrifuge the stock of Detection Reagent A and B briefly before use. Dilute to working concentration (1:100) with Assay Diluent A or B, respectively.
    4. Wash Solution - Dilute 20 mL of Wash Solution Concentrate (30x) with 580 mL of deionized or distilled water to make 600 mL of Wash Solution (1x).
    5. TMB Substrate - Aspirate the required amount of solution with sterile tip and do not return the residual solution back into the vial.

    Note:

    1. Serial dilution directly in the wells is not recommended.
    2. Prepare standard within 15 minutes before assay. Do not dissolve the reagents directly at 37 °C.
    3. Detection Reagent A and B are sticky solutions, so pipette them slowly to reduce volume errors.
    4. Reconstitute Standard or working solutions of Detection Reagent A and B carefully according to instructions, avoiding foaming and mixing gently until crystals are completely dissolved. To minimize inaccuracy caused by pipetting, use small volumes and ensure pipettes are calibrated. It is recommended to aspirate more than 10 µL for one-time pipetting.
    5. The reconstituted Standard, Detection Reagent A and B can only be used once.
    6. When crystals have formed in the Wash Solution concentrate (30x), warm it to room temperature and mix gently until the crystals are completely dissolved.
    7. Contaminated water or preparation containers affect the detection result.
    Aufbereitung der Proben
    • It is recommended to use fresh samples without long storage, otherwise protein degradation and denaturation may occur in these samples, leading to false results. Samples should therefore be stored for a short period at 2 - 8 °C or aliquoted at -20 °C (≤1 month) or -80 °C (≤ 3 months). Repeated freeze-thaw cycles should be avoided. Prior to assay, the frozen samples should be slowly thawed and centrifuged to remove precipitates.
    • If the sample type is not specified in the instructions, a preliminary test is necessary to determine compatibility with the kit.
    • If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibility of causing a deviation due to the introduced chemical substance. The recommended dilution factor is for reference only.
    • Please estimate the concentration of the samples before performing the test. If the values are not in the range of the standard curve, the optimal sample dilution for the particular experiment has to be determined.
    Testpräzision
    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level of target were tested 20 times on one plate, respectively.
    Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level of target were tested on 3 different plates, 8 replicates in each plate.
    CV(%) = SD/meanX100
    Intra-Assay: CV < 10%
    Inter-Assay: CV < 12%
    Beschränkungen
    Nur für Forschungszwecke einsetzbar
  • Lagerung
    4 °C/-20 ° C
    Informationen zur Lagerung
    1. For unopened kit: All reagents should be stored according to the labels on the vials. The Standard, Detection Reagent A, Detection Reagent B, and 96-well Strip Plate should be stored at -20 °C upon receipt, while the other reagents should be stored at 4 °C.
    2. For opened kits: the remaining reagents must be stored according to the above storage conditions. In addition, please return the unused wells to the foil pouch containing the desiccant and seal the foil pouch with the zipper.
    Haltbarkeit
    6 months
  • Target Alle APOA4 ELISA Kits anzeigen
    APOA4 (Apolipoprotein A-IV (APOA4))
    Andere Bezeichnung
    Apolipoprotein A4 (APOA4 Produkte)
    Synonyme
    Apo-AIV ELISA Kit, ApoA-IV ELISA Kit, apoAIV ELISA Kit, APOAIV ELISA Kit, APOA4 ELISA Kit, myh11 ELISA Kit, Apoa-4 ELISA Kit, apolipoprotein A4 ELISA Kit, apolipoprotein A4 L homeolog ELISA Kit, apolipoprotein A-IV ELISA Kit, apolipoprotein A-IV a ELISA Kit, APOA4 ELISA Kit, Apoa4 ELISA Kit, apoa4 ELISA Kit, apoa4.L ELISA Kit, apoa4a ELISA Kit
    Hintergrund
    ApoA-IV, ApoAIV, Apo-A4
    Pathways
    Lipid Metabolism